beautypg.com

4 simulating a transmitter, Loop power supply 30 vdc max uut – Martel Electronics MC1010 User Manual

Page 18

background image

18

4.3-1 HART™ Resistor Selection

The MC1010 can be set-up so that the 250 ohm resistor required for Hart™ configuration
devices resides inside the MC1010. Enabling the MC1010's internal 250 ohm resistor
eliminates the need to manually add a series resistor during a Hart™ calibration process.

NOTE: When the MC1010's internal 250 resistor is enabled, maximum load driving capability
drops from 1000 ohms @ 20mA to 750 ohms @20mA.

Enable/Disable Procedure

1. Remove the battery cover and remove the 2 screws that are at the top of the case.

2. Remove the 2 screws on the bottom or lower portion of the case.

3. Gently remove the top half of the case from the bottom.

4. Figure 12. shows the location of the Hart™ jumpers.

Figure 12.

4.4 Simulating a Transmitter

To have the calibrator supply a variable test current to a loop in place of a transmitter, follow
these steps:

1.

Select the [CONFIG] option from the main menu.

2.

Choose mA simulation from the primary parameters [mA 2W SIM], and enter the desired
current.

3.

Connect the 24V loop as shown in Figure 13.

Figure 13. Connections for Simulating a Transmitter

MEASURE / SOURCE

3W
mA+

TC

+

V

Hz

Ω

4W
mA–

+

+

LOOP
POWER
SUPPLY

30 VDC
MAX

UUT