For example:

If when measuring a P100-385 at 0 degrees cel sending OHMS? would return
1.000000E+02, OHM .

OUT

This command also switches the calibrator into mA output mode. A number and a
unit must be entered after the command.

For example:

OUT 5 MA sets the current output at 5 mA

OUT?

Returns the output of the calibrator.

Using the above example, OUT? Would return: 5.000000E-03, A

PRES_UNIT

Used to set the pressure unit for the indicated display

For example:

To set the pressure unit to psi on the lower display send PRES_UNIT LOWER,
PSI.

PRES_UNIT?

Returns the pressure unit used when measuring pressure for each of the 3 displays.

PUMP_LIMIT

Sets the approximate pressure in psi at which the pump will turn off.

For example:

PUMP_LIMIT 50 sets the approximate value that the pump will shutoff at to 50 psi

PUMP_LIMIT?

Returns the pump limit. Using the above example, PUMP_LIMIT? Would return:

50.000

REMOTE

Puts the calibrator in remote mode. From the remote mode the user can still use the
keypad to get back to local unless the command LOCKOUT was entered before
REMOTE. Then the keypad is totally locked out, and the user has to send the LOCAL
command to get back to local operation.

RTD_TYPE

Sets the RTD type. The following is a list of RTD types the way they should be
entered after the command: