Class I Test Flowchart
Start
#
CAUTION
• Short-circuit between L/N and PE is doubt when "no" and "OFF"
appears on the LCD in turns at the insulation resistance measurement
of the Extension leads test for Class I & II while DUT is on. Testing
should be immediately suspended.
• Follow the procedure described in 5.2-3 and do NULL setting before a
measurement.
• Alligator clip must make good contact with the enclosure of the DUT.
• When the terminal is open or the resistance value exceeds measuring
range, “OL” mark (over range display) appears on the LCD.
• Do not touch the device under test while testing is in progress. Since
a high voltage of 500V, user may get electrical shock.
⑴
. Protective conductor resistance.
RPE <
= 1Ω?
⑵
.Appliance switch test
(3). Insulation resistance between L/N and PE.
LnE >
= 1MΩ?
Light up in red
“
no” →“Con”→“value” will
be repeated on the LCD.
Light up in red
“
no” →“LnE”→“value” will
be repeated on the LCD.
Value (1). and (3). will be
alternately displayed on LCD
When the resistance between
LN is about 100kΩ or more
“
no” →“OFF” is displayed
alternately on the LCD.
No
No
No
Yes
Yes
Yes
PASS
Light up in green
Light up in green
Is the Switch