4. Specification
4.1 General specification, measuring range and accuracy
Voltage(VOLT) measurement of main power supply
Measuring range
207 ~ 264V AC
Resolution
1V
Accuracy
± (2%rdg+3dgt)
Measurement of Protective conductor resistance(RPE)
Measuring range
0 ~ 15.00Ω
Resolution
10mΩ
Open-circuit voltage
Measuring current 10A AC nominal value Accuracy ± (3%rdg+5dgt) Measurement of Insulation resistance (RINS) Rating 250V/20MΩ and 500V / 20MΩ Measuring range 0.1~19.99MΩ Resolution 10kΩ Rated voltage 250V/500V DC(+20%/-10%) @1MΩ Short-circuit current 2.5mA DC or less Accuracy ± (2% rdg+3dgt) Leakage current test Measuring range AC0.1~ 19.99mArms Resolution 0.01mA Accuracy ±(3%rdg ± 5dgt) Examination time Max 15 seconds Note: For MOV appliances use Leakage Current test. 4.2 Threshold and display Equipment Protective conductor resistance Insulation resistance Leakage current Class I RPE = < 1Ω RINS > = 1MΩ IEL = < 5mA Class II RINS > = 1MΩ IEL = < 1mA Extension Leads RPE = < 1Ω RINS > = 1MΩ IEL = < 1mA
- 1009 (13 pages)
- 1011 (12 pages)
- 1012 (12 pages)
- 1018 (1 page)
- 1030 (1 page)
- 1061 (58 pages)
- 1051 (40 pages)
- 1109S (36 pages)
- 1110 (1 page)
- 2000 (1 page)
- 2002PA (27 pages)
- 2007A (1 page)
- 2009R (33 pages)
- 2012R (2 pages)
- 2017 (1 page)
- 2031 (1 page)
- 2033 (9 pages)
- 2040 (2 pages)
- 2046R (2 pages)
- 2200 (3 pages)
- 2210R (1 page)
- 2300R (1 page)
- 2413F (24 pages)
- 2413R (24 pages)
- 2431 (2 pages)
- 2432 (1 page)
- 2433R (1 page)
- 2434 (10 pages)
- 2500 (1 page)
- 2608A (2 pages)
- 3005A (24 pages)
- 3021 (24 pages)
- 3121A (12 pages)
- 3126 (28 pages)
- 3127 (60 pages)
- 3128 (88 pages)
- 3131A (20 pages)
- 3132A (20 pages)
- 3161A (24 pages)
- 3321A (24 pages)
- 4105A (10 pages)
- 4106 (48 pages)
- 4116A (20 pages)
- 4140 (32 pages)
- 4200 (24 pages)