Specification, Measurement specification, Continuity – KYORITSU 6010B User Manual

Page 8: Loop impedance, Rcd uc, Insulation resistance

background image

+5% +15%rdg

±8dgt

5mA at IΔn=10mA

15mA at IΔn=30/100mA

150mA at IΔn=300/500mA

Test Current

Accuracy

Test Current

Trip Time

Open Circuit

Voltage (DC)

Short Circuit

Current

Range

Accuracy

Greater than 6 V

Greater than

200mA@2Ω

20/200Ω

Auto - Ranging

± (3%rdg + 4dgt)
± (3%rdg + 3dgt)

Up to 2Ω

Over 2Ω

― 6 ―

4. SPECIFICATION

Measurement Specification

Accuracy

Range

100V

Rated Voltage (AC)

230V+10%-15% 50Hz

@ KAMP10 Test lead

Continuity

Rated Voltage

(AC)

Nominal Test Current

at 0

Ω External Loop

Range

Accuracy

230V+10%-15%

50Hz

25A/10ms

15mA/350ms max.

± (3%rdg + 8dgt)

20Ω

2000Ω

Loop Impedance

Function

Rated

Voltage

(AC)

Test Current

Test

Current

Duration

x1/2

x1

FAST

DC

Auto

Ramp

230V

+10%

-15%

50Hz

10/30/100/300/500mA

10/30/100/300/500mA

150mA

10/30/100/300mA

500mA

Goes up by 10% from 20%

to 110% of IΔn. 300ms x 10

± (1%rdg

+3dgt)

-

2000ms

2000ms

50ms

2000ms

200ms

-8% -2%

+2% +8%

±10%

±4%

RCD

Uc

Function

Open Circuit

Voltage (DC)

Rated Current

Range

Accuracy

500V

1000V

500V+20%-0%

1000V+20%-0%

1mA or greater

@ 500kΩ

1mA or greater

@ 1MΩ

± (3%rdg + 3dgt)

20/200MΩ

Auto-Ranging

Insulation Resistance