Specification, Measurement specification, Continuity – KYORITSU 6010B User Manual
Page 8: Loop impedance, Rcd uc, Insulation resistance
+5% +15%rdg
±8dgt
5mA at IΔn=10mA
15mA at IΔn=30/100mA
150mA at IΔn=300/500mA
Test Current
Accuracy
Test Current
Trip Time
Open Circuit
Voltage (DC)
Short Circuit
Current
Range
Accuracy
Greater than 6 V
Greater than
200mA@2Ω
20/200Ω
Auto - Ranging
± (3%rdg + 4dgt)
± (3%rdg + 3dgt)
Up to 2Ω
Over 2Ω
― 6 ―
4. SPECIFICATION
Measurement Specification
Accuracy
Range
100V
Rated Voltage (AC)
230V+10%-15% 50Hz
@ KAMP10 Test lead
Continuity
Rated Voltage
(AC)
Nominal Test Current
at 0
Ω External Loop
Range
Accuracy
230V+10%-15%
50Hz
25A/10ms
15mA/350ms max.
± (3%rdg + 8dgt)
20Ω
2000Ω
Loop Impedance
Function
Rated
Voltage
(AC)
Test Current
Test
Current
Duration
x1/2
x1
FAST
DC
Auto
Ramp
230V
+10%
-15%
50Hz
10/30/100/300/500mA
10/30/100/300/500mA
150mA
10/30/100/300mA
500mA
Goes up by 10% from 20%
to 110% of IΔn. 300ms x 10
± (1%rdg
+3dgt)
-
2000ms
2000ms
50ms
2000ms
200ms
-8% -2%
+2% +8%
±10%
±4%
RCD
Uc
Function
Open Circuit
Voltage (DC)
Rated Current
Range
Accuracy
500V
1000V
500V+20%-0%
1000V+20%-0%
1mA or greater
@ 500kΩ
1mA or greater
@ 1MΩ
± (3%rdg + 3dgt)
20/200MΩ
Auto-Ranging
Insulation Resistance