KYORITSU 6010A User Manual

Page 9

background image

― 7 ―

To prevent wrong connection of test leads and to maintain safety, the
dedicated terminals used for continuity and insulation tests are automatically
covered when using the terminals for loop impedance and RCD tests

Instrument dimensions

175 X 115 X 86mm

Instrument weight

780g including batteries.

Reference conditions

Specifications are based on the following
conditions except where otherwise stated:-

1.Ambient temperature: 23± 5:
2.Relative humidity: 45% to 75%
3.Position: horizontal
4.AC power source: 230V, 50Hz
5.DC power source: 12.0 V, ripple content 1%

or less

6. Altitude: Up to 2000m

Battery type

Eight R6 or LR6 batteries.

Low battery warning

"

" symbol appears in the display if the battery

voltage drops below 8V.

Maximum numbers of

Approx. 700 times for use by continuity range or

measurements

approx. 1000 times for use by insulation resistance
range.

Operating temperature

0 to +40:, relative humidity 80% or less, no

and humidity.

condensation.

Storage temperature

-10 to +50:, relative humidity 75% or less, no

and humidity

condensation.
Illuminates if there is an alternating voltage of 20V
AC or more in the circuit under test before
continuity or insulation resistance tests. When DC
voltage is detected across the measuring terminal
the LED lights up.
The P-E and P-N LEDs illuminate when the wiring
of the circuit under test is correct. The

reverse

lamp is lit when P and N are reversed.

Auto data hold

In the loop impedance and RCD test functions, the
LCD reading is automatically frozen for 5 seconds
after measurement.

Display

The liquid crystal display has 3 1/2 digits with a
decimal point and units of measurement (Ω, MΩ,
V and ms) relative to selected function. The display

LED indication of live
circuit warning

LED indication of correct
Polarity