KYORITSU 6010A User Manual
Page 10
― 8 ―
is updated approximately five times per second.
Overload protection
The continuity test circuit is protected by a 0.5 A
600 V fast acting(HRC)ceramic fuse mounted in
the battery compartment, where a spare fuse is
also stored.
The insulation resistance test circuit is protected by
a resistor against 600 V AC for 10 seconds.
Mains Voltage Indication
On connecting test leads to the circuit under test
on Loop and RCD ranges, the LCD reads V-PE.
Sign "V-PE Lo" or "V-PE Hi" is also shown when
the voltage is 100V or less, or 260V or greater
respectively.
Equipment protected throughout by DOUBLE
INSULATION or REINFORCED INSULATION.
Caution, risk of electric shock.
Caution(refer to accompanying instruction
manual)
.
Symbols used on the
instrument
See also other documents in the category KYORITSU Tools:
- 1009 (13 pages)
- 1011 (12 pages)
- 1012 (12 pages)
- 1018 (1 page)
- 1030 (1 page)
- 1061 (58 pages)
- 1051 (40 pages)
- 1109S (36 pages)
- 1110 (1 page)
- 2000 (1 page)
- 2002PA (27 pages)
- 2007A (1 page)
- 2009R (33 pages)
- 2012R (2 pages)
- 2017 (1 page)
- 2031 (1 page)
- 2033 (9 pages)
- 2040 (2 pages)
- 2046R (2 pages)
- 2200 (3 pages)
- 2210R (1 page)
- 2300R (1 page)
- 2413F (24 pages)
- 2413R (24 pages)
- 2431 (2 pages)
- 2432 (1 page)
- 2433R (1 page)
- 2434 (10 pages)
- 2500 (1 page)
- 2608A (2 pages)
- 3005A (24 pages)
- 3021 (24 pages)
- 3121A (12 pages)
- 3126 (28 pages)
- 3127 (60 pages)
- 3128 (88 pages)
- 3131A (20 pages)
- 3132A (20 pages)
- 3161A (24 pages)
- 3321A (24 pages)
- 4105A (10 pages)
- 4106 (48 pages)
- 4116A (20 pages)
- 4140 (32 pages)
- 4200 (24 pages)