KYORITSU 2009R User Manual
Page 12
̶ 10 ̶
●Operating System
ΔΣ modulation
●Display
Liquid crystal display with a maximum
count of 4200 plus annunciators
●Overrange Indication
"OL" is shown on the display
●Response Time
Approx. 2 seconds.
●Sample Rate
About 3 times per second.
●Location for use
Indoor use, Altitude up to 2000m
●Temperature and Humidity 23 +/-5℃ , relative humidity up to 75%
for Guranteed Accuracy
without condensation
●Operating Temperature
0-40℃ , relative humidity up to 85%
and Humidity
without condensation
●Storage Temperature
-20-60℃ , relative humidity up to 85%
and Humidity
without condensation
●Power Source
Two 1.5VDC R6P (SUM-3) batteries
●Current Consumption
Approx. 40mA max (ACA).
●Auto-power-off function
Automatically powered down in about
10 minutes after the last switch
operation
(power consumption: about 200μA)
●Overload Protection
DC/AC current ranges:
2400A AC for 10sec
DC/AC voltage ranges:
1200V AC/DC for 10sec
Resistance range:
1000V AC/DC for 10sec
●Withstand Voltage
6880V AC, 50/60Hz for 5 seconds
between electrical circuit and housing
case or metal part of the jaws
●Insulation Resistance
10MΩ or greater at 1000V between
electrical circuit and housing case or
metal part of the jaws
●Conductor Size
Approx. 55㎜ diameter max.
●Dimensions
250(L)x105(W)x49(D)mm
●Weight
About 540g
●Accessories
Test leads M-7017
R6P battery
Carrying case M-9094
Instruction manual
Recorder output Plug M-8201
●Optional Accessories
Multi-Tran M-8008
Output Lead M-7014.