6. Battery and Fuse Replacement
W ARNIN G
Be careful not to burn yourself.
• Fuse becomes a high temperature after current measurement, it is dangerous by touching it
directly.
When fuse or batteries are replaced after current measurement, please be sure to leave the
main unit for 10 minutes for cooling.
6.1 Battery Replacement
If the batteries fall below the normal operating voltage, the “
”symbol turns on.
Follow the steps below to replace the batteries with new ones.
(AA-size (R6/LR6) 1.5V batteries)
W ARNIN G
• Be sure to disconnect the instrument from the circuit under test and test leads before
replacing the batteries.
• Turn the function switch to OFF (turn off the power) .
• Do not operate the instrument with the casing left open.
CAUTION
• Do not mix batteries of different types or new batteries with used ones.
• Make sure the polarities of the new batteries are exactly as shown on the battery holder.
To replace the batteries:
1) Remove the screw on the back of the
casing.
2) Remove the back cover.
3) Take the batteries out of the housing.
4) Replace the batteries with new ones.
5) Close the back cover and fasten it with
the screw.
Battery housing
Screw hole
Back cover removed
Blank cover
(For USB adapter,
Printer adapter)