Caution – KYORITSU 1052 User Manual
Page 23
22
4.3.9 Capacitor Measurement (
)
CAUTION
To avoid damage to instrument
• Turn off the power to the circuit under test before starting measurement in order to prevent
any excessive voltage from being applied to the instrument.
• Before starting measurement, be sure to discharge the capacitor under check.
1) Turn the function switch to the “
”position.
2) Plug the test leads into the input terminals.
3) Open the test lead and press the REL key in 10nF range to adjust the
capacitance to zero. (The display shows “ 0.00 ”.)
4) Connect the test leads to the circuit under test and then read the value
when it stabilizes.
Test leads
Red
Black
Note
The value (zero adjustment) remains displayed until power-off.
4.3.10 Frequency Measurement (Hz)
CAUTION
To avoid damage to instrument
Turn off the power to the circuit under test before starting measurement in order to prevent any
excessive voltage from being applied to the instrument.
1) Turn the function switch to the “
V” position.
2) Press the SELECT key to select the range of frequency. (The
display shows the unit of frequency.)
3) Plug the test leads into the input terminals.
4) Connect the test leads to the under test and then read the
value when it stabilizes.
Test leads
Red
Black