beautypg.com

2 accuracy – KYORITSU 1052 User Manual

Page 10

background image

9

3.2 Accuracy

Test conditions:
Temperature and humidity: 23±5ºC at 80%RH or less
Accuracy: ±(% of reading + digits)

Note: Each response time is a value to rated accuracy within selected range.

DC Voltage Measurement

V

Range

Resolution

Accuracy

Input Resistance

Maximum

Input Voltage

600mV

0.1mV

0.09+2

10MΩ

1000V

DC

1000V

rms AC

6V

0.001V

11MΩ

60V

0.01V

10MΩ

600V

0.1V

1000V

1V

0.15+2

NMRR:

60dB or more 50/60Hz±0.1%

CMRR:

120dB or more 50/60Hz (Rs=1kΩ)

Response time: 1 sec max.

AC Voltage Measurement

V

AC Coupling:

RMS value detection, sine wave

MEAN value detection and RMS value calibration (KEW1052 only)

Range

Resolution

Accuracy

Input

Impedance

Maximum

Input

Voltage

50/60Hz 40Hz to 500Hz 500Hz to 1kHz

600mV

0.1mV

0.5+5

1+5

1.5+5

10MΩ, <200pF

1000V

rms AC

1000V

DC

6V

0.001V

11MΩ, <50pF

60V

0.01V

10MΩ, <50pF

600V

0.1V

1000V

1V

-

Accuracy: At 5 to 100% of range and 1000V range is 200 to 1000V.

less than 1500V peak

For non-sinusoidal waveforms, add ±(2% + 2% of full scale), for Crest factor<3.

CMRR: 60dB or more DC to 60Hz (Rs=1kΩ)
4 counts or less is corrected to 0, Response time: 2 sec max.

DC Current Measurement

A

Range

Resolution

Accuracy

Voltage Drop

Maximum

Input Current

600μA

0.1μA

0.2+2

< 0.12mV/μA

440mA Protected by

a 440mA/1000V fuse.

6000μA

1μA

60mA

0.01mA

< 3.3mV/mA

440mA

0.1mA

0.5+5

6A

0.001A

< 0.1V/A

10A Protected by

a 10A/1000V fuse.

10A

0.01A

Response time: 1 sec max.

This manual is related to the following products: