3 measuring instructions, W arnin g – KYORITSU 1052 User Manual
Page 18
17
4.3 Measuring Instructions
W ARNIN G
To avoid damage to instrument or equipment
• Before starting measurement, make sure that the position of function switch and the input
terminals for connecting the test leads are appropriate for the desired mode of measurement.
• Temporarily remove the test leads from the device under test before operating the function
switch.
• Verify proper operation on a known source before use or taking action as a result of the
indication of the instrument.
Test leads here include a test leads with crocodile clip (optional accessory).
4.3.1 AC Voltage Measurement (
V,
mV)
1) Turn the function switch to the “
V ” or “mV ” position.
2) Press the SELECT Key when selecting the "mV".
( “
” is displayed.)
3) Plug the test leads into the input terminals.
4) Connect the test leads to the circuit under test and then read the value
when it stabilizes.
Test leads
Red
Black
4.3.2 DC Voltage Measurement (
V,
mV)
1) Turn the function switch to the “
V ”or “mV” position.
2) Plug the test leads into the input terminals.
3) Connect the test leads to the circuit under test and then read the value
when it stabilizes.
Test leads
Red
Black
Note
If “ mV ”range is selected and the test leads are left open-circuited, the instrument may
give a certain reading. This dose not affect your measurement.