Commissioning / operation – KROHNE OPTIWAVE 5200 PROFIBUS User Manual
Page 20
4
COMMISSIONING / OPERATION
20
OPTIWAVE 5200 C
www.krohne.com
10/2013 - 4003066901 - MA OPTIWAVE5200C-PA-R01-en-
86
3
Temperature Below -35°C
Measure the ambient temperature. De-energize the device until
the ambient temperature is back in the given range. If the
temperature does not stay in the correct range, make sure that
there is insulation around the signal converter.
87
3
Temperature Above +75°C
Measure the ambient temperature. De-energize the device until
the ambient temperature is back in the given range. If the
temperature does not stay in the correct range, make sure that
there is insulation around the signal converter.
90
3
Temperature out of range warning
Measure the ambient temperature. De-energize the device until
the ambient temperature is back in the given range. If the
temperature does not stay in the correct range, make sure that
there is insulation around the signal converter.
93
3
Local Operation
The device is in configuration mode. A supervisor has used the
LCD display of the device to enter a menu. Use the LCD display to
go back to normal mode.
100
3
No OW5200C Connection
De-energize the device. Make sure that the signal cable engages
in the terminal and the screw connection is tight. Energize the
device. If the problem continues, replace the signal converter. For
more data, refer to the handbook.
1 Indication is supported if Classic Diagnosis is active only.
2 Indication is supported if Condensed Diagnosis is active only.
3 Indication is supported in cyclic diagnosis only if Ident. Number 4554 hex is active. Nevertheless this information is always available via
acyclic access (e.g. by using DD / DTM) for all Ident. Numbers supported.
Bit Number
sup
p
orted
De
sc
ri
ption
Corrective action