BM 100 A technical data

4.1 Technical

data


Input
Measured variable

Distance, level, liquid interface and volume

Measurement range

Variable according to probe type, see Equipment architecture,
section 4.2

Blocking Distance

Variable according to probe type, see Probe measurement limits

Output
Analogue

4 … 20 mA (3.8 … 20.5 mA according to NAMUR 043),
1 passive output (as standard)

Load

Active outputs

350 ohms maximum

Passive outputs

(U-8)/22.10-3 ohms

Temperature drift

<100 ppm / K

Resolution

± 3 µA

Digital HART

®

& KROHNE SMART protocols (as standard).

PROFIBUS-PA output also available on demand.

Error signal

Status markers and text messages via local display or remote
interface. NAMUR 043-compliant (output current values at 3.6
mA or 22mA according to value configured in fct. 1.3.2).

Performance Characteristics
Accuracy

Liquids when probe length, L < 6 m / 20 ft.*
± 5 mm / ± 0.2”
Optimised with appropriate on-site calibration
± 3 mm / ± 0.12”
Liquids when probe length, L >6 m / 20 ft.*
Additional error above 6m: ±0.02% of distance measured

Level measurement

Solids (powders / granulates)**
± 20 mm / 0.79”

Interface measurement

Liquids (with minimum layer of 100 mm / 4”***
± 10 mm / 0.4” with configured and stable

ε

r

value

Repeatability

± 1 mm / ± 0.04”

Resolution

± 0.3 mm / ± 0.012”

Warm-up time

20 seconds to 1½ minutes (self-test before first reading)

* test medium: water (

ε

r

=80) ** test medium: cement (

ε

r

=3) *** test medium: oil (

ε

r

=2.4)

Test conditions
Ambient temperature

+20°C ±5°C (+70°F ±10°F)

Ambient air pressure

1013 mbar abs. ±20 mbar / 14.69 psig ±0.29 psi

Relative air humidity

60% ±15%

Reference target

coaxial probe: water surface
single probe: Ø0.8m / Ø31.5” metal plate
twin probe: metal block “short circuit”

Distance from wall

> 300 mm / 11.81” (not for coaxial probe)

Distance to obstruction

> 1 m / 3.28”


Environment
Ambient temperature

-20 … + 50°C / -4 … 120°F

Storage temperature

-40 … + 85°C / -40 … 185°F

Protection

IP 67 (US equivalent: NEMA 6 – 6P)