3

INSTALLATION

20

OPTIMASS 1000

www.krohne.com

04/2011 - 4000977901 - MA OPTIMASS 1000 R01 en

3.4.3 Cross talk

Where more than one meter is being installed, a very high level of immunity to cross talk means
that the meters can be mounted within close proximity to each other. The meters can be
mounted either in series or parallel, as shown.

3.4.4 Flange connections

Tighten the flange bolts evenly and in turn.

Meters in series

INFORMATION!
Where the meters are installed in series, it is strongly recommended that the process pipe
diameter remains constant. For more information, please contact the manufacturer.

Meters in parallel