6

APPLICATION SEQUENCES

24

UFC 400

www.krohne.com

05/2013 - 4002744801 - AD Modbus UFC 400 R01 en

6.2 Zero Calibration

• The example assumes slave address 1 for Modbus telegram.

1

Start

2

Start Zero Calibration

2a

Set Coil 2000 (0x07D0)
MB-Telegram: 01 05 07 D0 FF 00 + CRC

3

Check Status Of Zero Calibration

3a

Read Coil 2000 (0x07D0)
MB-Telegram: 01 01 07 D0 00 01 + CRC

4a

running

4b

ready

5
5a

Read Result
Read Input Register 20000 (0x4E20)
MB-Telegram: 01 04 4E 20 00 01 + CRC

6a

NOK

6b

OK

7

Calibration Failure

8

Apply Changes

8a

Set Coil 1002 (0x03EA)
MB-Telegram: 01 05 03 EA FF 00 + CRC

9

Calibration Done