Modbus Function Codes "Read Holding Registers" (0x03) and "Write Multiple Registers" (0x10)

Holding
Register

Name

Action

Type

No. of
registers

Values

Display
Fct. No.

40000

40000

40000

40000

(0x9C40)

Totaliser 1

Function

Byte

1

0 = Off

1 = Absolute total

2 = Increment

3 = Decrement

C3.1.1

40001

40001

40001

40001

(0x9C41)

Measurement

Byte

1

0 = Volume Flow

2 = Mass Flow

C3.1.2

40002

40002

40002

40002

(0x9C42)

Totaliser 2

Function

Byte

1

0 = Off

1 = Absolute total

2 = Increment

3 = Decrement

C3.2.1

40003

40003

40003

40003

(0x9C43)

Measurement

Byte

1

0 = Volume Flow

2 = Mass Flow

C3.2.2

40004

40004

40004

40004

(0x9C44)

Totaliser 3

1

Function

Byte

1

0 = Off

1 = Absolute total

2 = Increment

3 = Decrement

C3.3.1

40005

40005

40005

40005

(0x9C45)

Measurement

Byte

1

0 = Volume Flow

2 = Mass Flow

C3.3.2

1 Only available in signal converters with "modular carrier"

Modbus Function Codes "Read Holding Registers" (0x03) and "Write Multiple Registers" (0x10)

Holding
Register

Name

Action

Type

No. of
registers

Units

Display
Fct. No.

40500

40500

40500

40500

(0x9E34)

Totaliser 1

Low Flow Cut-Off

Value

Float

2

[m³/s] or [kg/s]

C3.1.3

40502

40502

40502

40502

(0x9E36)

Time Constant

0...100 [s]

C3.1.4

40504

40504

40504

40504

(0x9E38)

Set or Read Value

[m³] or [kg]

C3.1.7

40506

40506

40506

40506

(0x9E3A)

Preset

[m³] or [kg]

C3.1.5

40508

40508

40508

40508

(0x9E3C)

Totaliser 2

Low Flow Cut-Off

Value

Float

2

[m³/s] or [kg/s]

C3.2.3

40510

40510

40510

40510

(0x9E3E)

Time Constant

0...100 [s]

C3.2.4

40512

40512

40512

40512

(0x9E40)

Set or Read Value

[m³] or [kg]

C3.2.7

40514

40514

40514

40514

(0x9E42)

Preset

[m³] or [kg]

C3.2.5

40516

40516

40516

40516

(0x9E44)

Totaliser 3

1

Low Flow Cut-Off

Value

Float

2

[m³/s] or [kg/s]

C3.3.3

40518

40518

40518

40518

(0x9E46)

Time Constant

0...100 [s]

C3.3.4

40520

40520

40520

40520

(0x9E48)

Set or Read Value

[m³] or [kg]

C3.3.7

40522

40522

40522

40522

(0x9E4A)

Preset

[m³] or [kg]

C3.3.5