MODBUS PROTOCOL

5

19

UFC 400

www.krohne.com

05/2013 - 4002744801 - AD Modbus UFC 400 R01 en

5.8.7 Zero Calibration

A zero calibration is not necessary for normal operation. It's recommended that only trained
service engineers should perform a zero calibration.

The Coils and Registers below are required to perform a zero calibration. For further
information refer to

Zero Calibration

on page 24.

Modbus Function Codes "Read Coils" (0x01) and "Write Single Coil" (0x05)

Coil
Address

Name

Function

Action

Values

Display
Fct. No.

2000

2000

2000

2000

(0x07D0)

Zero Calibration

Control

Start

Write

1 = start zero calibration

C1.2.1

Status

Read

0 = zero calibration not running /

complete

1 = zero calibration in progress

-

Modbus Function Code "Read Input Registers" (0x04)

Input
Register

Name

Type

No. of
registers

Values

20000

20000

20000

20000

(0x4E20)

Zero Calibration Result

Byte

1

1 = OK

0 = NOK