TECHNICAL DATA

6

29

OPTIFLUX 1000

www.krohne.com

04/2014 - 4000846303 - HB OPTIFLUX 1000 R03 en

6.3 Measuring accuracy

Each flowmeter is standard wet calibrated under reference conditions by direct volume
comparison. The performance of the flowmeter is defined and documented in an individual
calibration certificate.

Reference conditions

• Medium: water
• Temperature: +10...30°C / +50...86°F
• Pressure: 1 bar / 14.5 psi
• Inlet section: ≥ 5 DN
• Electrical conductivity: ≥ 300 μS/cm

Accuracy

Figure 6-2: Flow velocity vs. accuracy

X [m/s]: flow velocity
Y [%]: deviation from the actual measured value (mv)

Sensor diameter

Converter
type

Accuracy

Curve

DN10...150 / 3/8...6"

IFC 050

0.5% of mv above 0.5 m/s

below 0.5 m/s, deviation ± 2.5 mm/s

3

DN10...150 / 3/8...6"

IFC 100

0.4% of mv + 1 mm/s

2

DN10...150 / 3/8...6"

IFC 300

0.3% of mv + 2 mm/s

1