ELECTRICAL CONNECTIONS

4

21

OPTIFLUX 1000

www.krohne.com

04/2014 - 4000846303 - HB OPTIFLUX 1000 R03 en

Grounding ring number 1 (optional for DN25...150):

• Thickness: 3 mm / 0.1" (tantalum: 0.5 mm / 0.02")

Note:

Note:

Note:

Note: For diameter DN10 and DN15, grounding rings are integrated as standard in the flow
sensor construction.

4.3 Virtual reference for IFC 300 (C, W and F version)

The virtual reference option on the IFC 300 flow converter provides complete isolation of the
measurement circuit.

Benefits of virtual reference:

• Grounding rings or grounding electrodes can be omitted.
• Safety increases by reducing the number of potential leakage points.
• The installation of the flowmeters is much easier.

Minimum requirements:

• Size: ≥ DN10
• Electrical conductivity: ≥ 200 µS/cm
• Electrode cable: max. 50 m / 164 ft, type DS

4.4 Connection diagrams

Figure 4-2: Grounding ring number 1

Figure 4-3: Virtual reference

INFORMATION!
For the connection diagrams, please refer to the documentation of the applicable converter.