6.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Measuring principle
Faraday's law of induction
Application range
Electrically conductive fluids
Measured value
Measured value
Measured value
Measured value
Primary measured value
Flow velocity
Secondary measured value
Volume flow
Design
Features
Sandwich (wafer) design
PFA liner and Hastelloy
®
electrodes
Light weight and compact
Modular construction
The measurement system consists of a flow sensor and a signal converter.
It is available as compact and as separate version. Additional information
can be found in the documentation of the signal converter.
Compact version
With IFC 050 converter: OPTIFLUX 1050 C
With IFC 100 converter: OPTIFLUX 1100 C
With IFC 300 converter: OPTIFLUX 1300 C
Remote version
In wall (W) mount version with IFC 050 converter: OPTIFLUX 1050 W
In wall (W) mount version with IFC 100 converter: OPTIFLUX 1100 W
In field (F), wall (W) or rack (R) mount version with IFC 300 converter:
OPTIFLUX 1300 F, W or R
Nominal diameter
DN10...150 / 3/8...6"
Measuring accuracy
Reference conditions
Medium: water
Temperature +10...+30°C / +50...+86°F
Operating pressure: 1 bar / 14.5 psi
Inlet section ≥ 5 DN
Electrical conductivity: ≥ 300 µS/cm
Maximum measuring error
IFC 050: down to 0.5% of the measured value above 0.5 m/s
below 0.5 m/s; deviation ± 2.5 mm/s
IFC 100: down to 0.4% of the measured value ± 1 mm/s
IFC 300: down to 0.3% of the measured value ± 2 mm/s
The maximum measuring error depends on the installation conditions.
For detailed information refer to
Measuring accuracy
on page 29.
Repeatability
±
0.1% of MV, minimum 1 mm/s
Calibration
2 point calibration by direct volume comparison.