KROHNE OPTISONIC V6 Modbus User Manual
Page 21

2011-07-19
OPTISONIC_V6_Modbus_6.doc
21/29
K
R
O
H
N
E
C
T
P
ro
d
u
c
ts
Parameter / Variable
(short description)
Name
Explanation
Relative
address
Absolute
address,
non-
Modicon
Absolute
address,
Modicon-
comp.
Units
Date and time
ANSI time
Built-in real time clock: date and time according to ANSI
standard: number of seconds elapsed since January 1
st
, 1970,
00:00:00 hours.
Note: Although this is a Holding Register, this register is read-
only. Use the KROHNE Flowmeter Configuration and
Monitoring software to set or adjust the real-time clock.
1
5501
5502-5503
(2 regs)
s
8.5
Input Registers (read-only): Double (64-bit floating-point), address range 6000-6499
Parameter / Variable
(short description)
Name
Explanation
Relative
address
Absolute
address,
non-
Modicon
Absolute
address,
Modicon-
comp.
Units
Test register
TestRegister
double64
Register reserved for testing communications and protocol
handling with this type of register, without affecting the operation
of the flow meter.
0
6000
6000-6003
(4 regs)
-
8.6
Holding Registers (read/write): Double (64-bit floating-point), address range 6500-6999
Parameter / Variable
(short description)
Name
Explanation
Relative
address
Absolute
address,
non-
Modicon
Absolute
address,
Modicon-
comp.
Units
Test register
TestRegister
double64
Register reserved for testing communications and protocol
handling with this type of register, without affecting the operation
of the flow meter.
0
6500
6500-6503
(4 regs)
-