7

DEFAULT SETTINGS

By means of a number of parameters the Modbus communication link can be adjusted to one’s needs or
preferences. When the OPTISONIC V6 meter is delivered these parameters are set to default values as
listed below:

Port 0:

Baud rate:

19200

Data bits:

8

Stop bits:

1

Parity:

Even

Modbus Mode:

RTU

Modbus End Code:

2

Addressing Mode:

Modicon compatible

Representation Mode:

Normal

Modbus Address:

237

Modbus Start Gap:

40 (bits)

Modbus End Gap:

20 (bits)

Modbus Start Code:

“:” (colon character)

Modbus End Code1:

ASCII 13 (carriage return, CR)

Modbus End Code2:

ASCII 10 (line feed, LF)

Modbus Time Out:

1 second


Port 1:

Baud rate:

115200

Data bits:

8

Stop bits:

1

Parity:

Even

Modbus Mode:

RTU

Modbus End Code:

2

Addressing Mode:

Modicon compatible

Representation Mode:

Normal

Modbus Address:

237

Modbus Start Gap:

40 (bits)

Modbus End Gap:

20 (bits)

Modbus Start Code:

“:” (colon character)

Modbus End Code1:

ASCII 13 (carriage return, CR)

Modbus End Code2:

ASCII 10 (line feed, LF)

Modbus Time Out:

1 second