8

SERVICE

22

OPTIFLEX 2200 C

www.krohne.com

01/2013 - 4000738603 - AD SIL OPTIFLEX 2200 R03 en

8.5 Calibration procedure

8.5.1 General notes

If the probe is not in the same customer order as the signal converter, it is necessary to calibrate
the device.

Equipment needed

• A device with the integrated display option.
• Process measurement and device configuration software (e.g. the DTM for PACTware™), if

the device does not have the integrated display option.

• An ammeter.
• Reference device: an approved level meter or indicator

Requirements

• The ammeter must be correctly calibrated and connected directly to device.
• You must have 2 reference points (levels) in the tank given by a different measurement

solution. These points are identified as "ref_pt1" and ref_pt2".

CAUTION!
Do not connect the ammeter across the safety system PLC because this configuration can open
the circuit breaker.
Make sure that the 2 reference points are not in the top or bottom dead zones. The dimensions of
these zones are given in the "Technical data" chapter (Measurement limits) of the Handbook.