Flange Lost Warning
Risk of device failure. Replace electronics module or contact the
supplier.
82
3
Snapshot Invalid
Do a check of the device, tank and the process.
Do a new snapshot recording. If necessary, contact the supplier.
84
3
Reference Position Outside Range
Risk of device failure. Replace electronics module or contact the
supplier.
85
3
Audio Signal Offset Outside Range
Risk of device failure. Replace electronics module or contact the
supplier.
86
3
Temperature Below -35°C
Measure the ambient temperature. De-energize the device until
the ambient temperature is back in the given range. If the
temperature does not stay in the correct range, make sure that
there is insulation around the signal converter.
87
3
Temperature Above +75°C
Measure the ambient temperature. De-energize the device until
the ambient temperature is back in the given range. If the
temperature does not stay in the correct range, make sure that
there is insulation around the signal converter.
90
3
Temperature out of range warning
Measure the ambient temperature. De-energize the device until
the ambient temperature is back in the given range. If the
temperature does not stay in the correct range, make sure that
there is insulation around the signal converter.
93
3
Local Operation
The device is in configuration mode. A supervisor has used the
LCD display of the device to enter a menu. Use the LCD display to
go back to normal mode.
100
3
No OF2200C Connection
De-energize the device. Make sure that the signal cable engages
in the terminal and the screw connection is tight. Energize the
device. If the problem continues, replace the signal converter. For
more data, refer to the handbook.
1 Indication is supported if Classic Diagnosis is active only.
2 Indication is supported if Condensed Diagnosis is active only.
3 Indication is supported in cyclic diagnosis only if Ident. Number 4550 hex is active. Nevertheless this information is always available via
acyclic access (e.g. by using DD / DTM) for all Ident. Numbers supported.
Bi
t Nu
mbe
r
supported
Description
Correction action