COMMISSIONING / OPERATION

20

OPTIFLEX 2200 C

www.krohne.com

10/2013 - 4003066801 - MA OPTIFLEX2200C-PA-R01-en-

80

3

Flange Lost Warning

Risk of device failure. Replace electronics module or contact the

supplier.

82

3

Snapshot Invalid

Do a check of the device, tank and the process.

Do a new snapshot recording. If necessary, contact the supplier.

84

3

Reference Position Outside Range

Risk of device failure. Replace electronics module or contact the

supplier.

85

3

Audio Signal Offset Outside Range

Risk of device failure. Replace electronics module or contact the

supplier.

86

3

Temperature Below -35°C

Measure the ambient temperature. De-energize the device until

the ambient temperature is back in the given range. If the

temperature does not stay in the correct range, make sure that

there is insulation around the signal converter.

87

3

Temperature Above +75°C

Measure the ambient temperature. De-energize the device until

the ambient temperature is back in the given range. If the

temperature does not stay in the correct range, make sure that

there is insulation around the signal converter.

90

3

Temperature out of range warning

Measure the ambient temperature. De-energize the device until

the ambient temperature is back in the given range. If the

temperature does not stay in the correct range, make sure that

there is insulation around the signal converter.

93

3

Local Operation

The device is in configuration mode. A supervisor has used the

LCD display of the device to enter a menu. Use the LCD display to

go back to normal mode.

100

3

No OF2200C Connection

De-energize the device. Make sure that the signal cable engages

in the terminal and the screw connection is tight. Energize the

device. If the problem continues, replace the signal converter. For

more data, refer to the handbook.

1 Indication is supported if Classic Diagnosis is active only.
2 Indication is supported if Condensed Diagnosis is active only.
3 Indication is supported in cyclic diagnosis only if Ident. Number 4550 hex is active. Nevertheless this information is always available via

acyclic access (e.g. by using DD / DTM) for all Ident. Numbers supported.

Bi

t Nu

mbe

r

supported

Description

Correction action