Commissioning / operation – KROHNE OPTIWAVE 5200 PROFIBUS User Manual
Page 19
COMMISSIONING / OPERATION
4
19
OPTIFLEX 2200 C
www.krohne.com
10/2013 - 4003066801 - MA OPTIFLEX2200C-PA-R01-en-
58
3
Converter ROM Error
Replace the signal converter.For more data, refer to the
handbook.
59
3
Sensor EEPROM Error
Replace the signal converter.For more data, refer to the
handbook.
60
3
Sensor RAM Error
Replace the signal converter.For more data, refer to the
handbook.
61
3
Sensor ROM Error
Replace the signal converter.For more data, refer to the
handbook.
63
3
Oscillator Frequency Failure
Replace the signal converter.For more data, refer to the
handbook.
64
3
Converter Voltage Error
Replace the signal converter.For more data, refer to the
handbook.
65
3
Sensor Voltage Error
Do a check of the power supply at the device terminals. Make sure
that voltage values are in the specified limits in menu item 2.2.2
DIAGNOSTIC (Configuration mode / Supervisor menu). If the
voltage is correct, replace the signal converter. For more data on
how to replace the signal converter, refer to the handbook.
66
3
Internal Communication Error
De-energize the device. Make sure that the signal cable engages
in the terminal and the screw connection is tight. Energize the
device. If the problem continues, replace the signal converter. For
more data, refer to the handbook.
67
3
Temperature Out of Range Error
Measure the ambient temperature. De-energize the device until
the ambient temperature is back in the given range. If the
temperature does not stay in the correct range, make sure that
there is insulation around the signal converter. If this error
occurs 2 times, replace the device.
68
3
Sensor not compatible
Go to menu 1.1.0 IDENT. in Configuration mode. Record the
version numbers of the device software given in menu items 1.1.2,
1.1.3 and 1.1.4. Give this data to the supplier.
72
3
Reference Lost Error
Speak to your supplier to make sure that the electronics are still
functioning correctly. Make sure that your installation has ESD
protection. For more data, refer to the handbook.
73
3
Level Lost Error
Measure the level of the contents in the tank using another
method of measurement. If the tank is empty (the level is below
the end of the probe), then fill the tank until product level is in the
measurement range. If the tank is full (the level is in the blocking
distance), then remove the contents from the tank until the level
is back in the measuring range. If the product was lost and the
tank is neither full nor empty, wait for the device to find the level
again.
If the device has to measure a product with εr≥1.6, refer to MEAS.
AMP. (Measured Pulse Amplitude, menu item 2.5.6) and then
adjust MEAS. THRESH. (measurement threshold, menu item
2.5.7). If the product has a low dielectric constant (εr<1.6) and the
device is in TBF mode, refer to PROBE END AMP (probe end pulse
amplitude, menu item 2.5.8) and then adjust PROBE END TH.
(probe end threshold, menu item 2.5.9). For more data, refer to
the handbook.
Make sure that the signal converter is correctly attached the
probe. For more data, refer to the handbook.
75
3
Overfill Error
Remove some of the product until the level is below the blocking
distance.
76
3
No Probe detected Error
Replace the signal converter.For more data, refer to the
handbook.