Modbus protocol, 3 addressing, 4 overview of supported functions – KROHNE MFC 300 Modbus User Manual
Page 9
MODBUS PROTOCOL
6
9
MFC 300
www.krohne.com
11/2009 - 4000744801 - AD Modbus MFC 300 R01 en
6.3 Addressing
In the following tables the Modbus protocol addresses / data addresses are listed.
Some systems cannot use addresses above 9999. For these systems there is the possibility to
use the listed addresses but
• for Input Registers omit the leading 3 of 3xxxx;
• for Holding Registers omit the leading 4 of 4xxxx;
• for Input Registers replace the leading 20 of 20xxx by 9xxx.
Sometimes register numbers are asked for. The register numbers
register numbers
register numbers
register numbers can be calculated by adding a
1 to the protocol address and using a prefix according to the block:
• prefix 1 for coils
• prefix 3 for Input Registers
• prefix 4 for Holding Registers
6.4 Overview of supported functions
The following table shows Modbus functions supported by RS485 interface.
Function code
Name
Access to
hex
dec
01
01
Read Single Coil
Status of calibration functions, counter
status (start/stop)
03
03
Read Holding Register
Converter configuration parameter
04
04
Read Input Register
Measurement values, status values and
calibration results
05
05
Write Single Coil
Cold start, warm start, error reset, start
calibration function, start/stop counter
08
08
Diagnostics
-
10
16
Write Multiple Register
Converter configuration parameter