Modbus protocol – KROHNE MFC 300 Modbus User Manual
Page 11
MODBUS PROTOCOL
6
11
MFC 300
www.krohne.com
11/2009 - 4000744801 - AD Modbus MFC 300 R01 en
6.6.2 Counter controls
6.6.3 Start calibration functions
Modbus protocol
address
Description
Settings
Converter Fct. No.
3000
Start / Stop
Counter 1
Write
1
start counter
C4.1.8 / C4.1.9
Write
0
stop counter
Read
1
counter is running
Read
0
counter is stopped
3001
Start / Stop
Counter 2
Write
1
start counter
C4.2.8 / C4.2.9
Write
0
stop counter
Read
1
counter is running
Read
0
counter is stopped
3002
Start / Stop
Counter 3
1
Write
1
start counter
C4.3.8 / C4.3.9
Write
0
stop counter
Read
1
counter is running
Read
0
counter is stopped
3003
Reset Counter 1
Write
1
reset counter
C4.1.6
Write
0
-
Read
0
-
3004
Reset Counter 2
Write
1
reset counter
C4.2.6
Write
0
-
Read
0
-
3005
Reset Counter 3
1
Write
1
reset counter
C4.3.6
Write
0
-
Read
0
-
1 Only available in converters with IO2. A write attempt to a non-existing counter will cause an error response.
Modbus protocol
address
Description
Settings
Converter Fct. No.
2000
Zero Calibration
Write
1
start function
C1.1.1
Write
0
-
Read
0
-
2001
Single Point
Density Calibration
Write
1
start function
C1.2.1
Write
0
-
Read
0
-
2002
Two Point Density
Calibration - first
calibration point
Write
1
start function
C1.2.1
Write
0
-
Read
0
-
2003
Two Point Density
Calibration -
second calibration
point
Write
1
start function
C1.2.1
Write
0
-
Read
0
-