beautypg.com

Local configuration – KROHNE MFC 300 Modbus User Manual

Page 6

background image

4

LOCAL CONFIGURATION

6

MFC 300

www.krohne.com

11/2009 - 4000744801 - AD Modbus MFC 300 R01 en

Local configuration

Converter Fct.
No.

Display

Description and settings

C6.8.1

Slave Address

Selects the Modbus address of the device.

Range: 1..247 (default = 1)

C6.8.2

Baud Rate

Selects the baud rate of the device.

Options:

1200 / 2400 / 3600 / 4800 / 9600 / 19200 (default) / 38400 / 57600 / 115200

C6.8.3

Parity

Selects the parity.

Options:

Even (default) / Odd / No

C6.8.4

Data Format

Selects the data format.

Options:

Big Endian (default) / Little Endian

C6.8.5

Transmission Delay

Selects the delay between receiving the last byte of a request and

sending the first byte of the response.

Range: 0..40ms (default = 0ms)

C6.8.6

Stop Bits

Selects the number of stop bits.

Options:

1 (default) / 2

C6.8.7

Information

Displays information about the device.

.book Page 6 Tuesday, November 24, 2009 1:48 PM