Modbus protocol, 8 holding registers – KROHNE MFC 300 Modbus User Manual

Page 13

background image

MODBUS PROTOCOL

6

13

MFC 300

www.krohne.com

11/2009 - 4000744801 - AD Modbus MFC 300 R01 en

6.8 Holding registers

Some parameters of the device can be accessed as Modbus holding registers.

Function code 03 (0x03) for "Read" operations and function code 16 (0x10) for "Write" operations.

The holding registers are grouped into the following different sections.

6.8.1 Counter parameters

INFORMATION!
Counter 3 parameters are only available for converter with IO2.

Modbus protocol
address

Description and settings

Converter
Fct. No.

Type

Number of
registers

40000

Counter 1 Function

Counter 1 Function

Counter 1 Function

Counter 1 Function

1 = Sum Counter

2 = + Counter

3 = - Counter

0 = Off

C4.1.1

byte

1

40001

Measurement Counter 1

Measurement Counter 1

Measurement Counter 1

Measurement Counter 1

21 = Volume Flow

22 = Mass Flow

27 = Concentration Flow 1

28 = Concentration Flow 2

C4.1.2

word

1

40002

Counter 2 Function

Counter 2 Function

Counter 2 Function

Counter 2 Function

1 = Sum Counter

2 = + Counter

3 = - Counter

0 = Off

C4.2.1

byte

1

40003

Measurement Counter 2

Measurement Counter 2

Measurement Counter 2

Measurement Counter 2

21 = Volume Flow

22 = Mass Flow

27 = Concentration Flow 1

28 = Concentration Flow 2

C4.2.2

word

1

40004

Counter 3 Function

Counter 3 Function

Counter 3 Function

Counter 3 Function

1 = Sum Counter

2 = + Counter

3 = - Counter

0 = Off

C4.3.1

byte

1

40005

Measurement Counter 3

Measurement Counter 3

Measurement Counter 3

Measurement Counter 3

21 = Volume Flow

22 = Mass Flow

27 = Concentration Flow 1

28 = Concentration Flow 2

C4.3.2

word

1

41000

Low Flow Cutoff Value Counter 1

Low Flow Cutoff Value Counter 1

Low Flow Cutoff Value Counter 1

Low Flow Cutoff Value Counter 1 [m

3

/s]

or [kg/s]

C4.1.3

float

2

41002

Time Constant Counter 1

Time Constant Counter 1

Time Constant Counter 1

Time Constant Counter 1 [s]

C4.1.4

float

2

41004

Set Counter 1 or Read Counter 1

Set Counter 1 or Read Counter 1

Set Counter 1 or Read Counter 1

Set Counter 1 or Read Counter 1 [m

3

] or [kg]

C4.1.7

float

2

41006

Low Flow Cutoff Value Counter 2

Low Flow Cutoff Value Counter 2

Low Flow Cutoff Value Counter 2

Low Flow Cutoff Value Counter 2 [m

3

/s]

or [kg/s]

C4.2.3

float

2

41008

Time Constant Counter 2

Time Constant Counter 2

Time Constant Counter 2

Time Constant Counter 2 [s]

C4.2.4

float

2

41010

Set Counter 2 or Read Counter 2

Set Counter 2 or Read Counter 2

Set Counter 2 or Read Counter 2

Set Counter 2 or Read Counter 2 [m

3

] or [kg]

C4.2.7

float

2

.book Page 13 Tuesday, November 24, 2009 1:48 PM