Manual OPTISENS PAC 050

OPTISENS PAC 050 24


6.1 Error messages

Error message

Cause

Action

Slope error

The slope determined by

calibration was higher than

65 mV/pH or lower than 50 mV/pH.

Please check the sensor connection and

cable, and make sure that the calibration data

stored in the instrument corresponds to the

calibration solutions used. Also make sure that
the temperature settings are correct. Then

repeat the calibration. If the error message

remains, the sensor has to be replaced or

regenerated.

Sensor zero

The zero deviation determined by

calibration was higher than 60 mV

or lower than -60 mV.

Please follow the procedure described above.

Make sure that the sensor buffer stored in the

calibration data corresponds to the sensor
used. If the error remains, at least the

reference electrode has to be replaced or

regenerated.

Error input 1

The pH or ORP electrode gives no

signal.

Please check the connections, the cable and

the sensor for signs of damage.

Error input 2

The temperature sensor gives no

signal.

Please check the connections, the cable and

the sensor for signs of damage.

This message also appears if automatic

temperature compensation was selected
although no temperature sensor was used.

Limit 1 / 2

The measured value exceeded

limit 1 (or dropped below limit 2,
respectively).

Please check the dosing and readjust the

control parameters, if necessary.

Dosage

Pump 1 (or 2, resp.) feeds with

100% for more than the defined
period of time.

Please check the dosing, especially the

feeding tubes and connections. Caution!
Carefully check for leaking chemicals!

Ext. controller stop

The digital input has been short-

circuited.

This only indicates the external controller stop.

If, however, you have connected a level
sensor, this message appears due to the „lack

of water“ alarm.

7. Output

Main menu

.

.

Basic settings

Basic settings

.

.

Analog output

Analog

output

0…20mA/4…20mA

.

Start

0/4mA

.

End 20 mA

.

Analog

output

Meas.

value

Serial interface
RS485 (option)

Bus

address

Bus address

Nr.

00

Set a number
between 0 and 31 to

address the device

via data bus.