4.5 Temperature compensation

Main menu

Automatic

compensation

Manual

compensation

.

Temp. comp.

Temp. comp.

Temp. comp.

Auto comp.

Manual comp.

.

Temperature

.

25.0°C

.

Basic settings

Calibration
temperature

Basic settings

Cal. temp.

Cal. temp.

0.0°C

Select between two ways of compensation:

1) Automatic compensation with temperature sensor

Mind that the temperature sensor should always measure the medium temperature around the sensor. When
pH sensor and temperature sensor are not immersed in the same solution, better switch to manual

compensation.

2) Manual compensation
If the temperature can be regarded as constant, you can enter the temperature manually instead of

measuring it continuously. The device will then compensate the temperature error of this temperature.

4.5.1 Calibration of the temperature measurement


If you connect the temperature sensor with a twin-core cable, slight deviations might occur between

measured and real temperature. These deviations can be eliminated by calibration.

During start-up, measure the temperature manually and enter a correction term so that the display shows the
exact temperature.

5. Adjustments of the controller


Main menu

.

.

Selection of ON/OFF, P or PI controller, adjustment of P ranges and I

Set points

functions for 2 set points

.

.

Basic settings

Basic settings

.

.

Contr.

settings

Adjustment of frequencies and pulse/pause

times, hysteresis, direction of the control

action, delay.

For any type of controller you have to enter one or two set points, and you have to tell the device whether

these set points are reached by increasing or decreasing the measured value.