3.3 Password and language

Main menu

.

.

Enter password

Enter password

056

Code

3.3.1 Enter password

To get access to the various parameters you have to enter the correct password:
• Code 11 gives access to the parameters „calibration“, „temperature compensation“, and „set points“.
• Code 86 gives access to all parameters and functions.
With any other number it is impossible to select, view or change any parameter.

Main menu

.

.

Basic settings

Basic settings

.

.

Language

Language

We speak

English

3.3.2 Language

For the communication with the device you can choose from a variety of languages.

Since choosing a language is part of the basic settings, it requires code 86. If a different code is set, you will

be asked to enter the correct password.

4. Adjustments for the measurement


Main menu

Calibrate

Calibration of the pH measurement, display of slope and zero-point.

Temp. comp.

.

Selection between automatic and manual temperature compensation,

adjustment of temperature for manual compensation.

.

.

Basic settings

Basic settings

Cal. Temp.

Calibration of the temperature measurement.

Buffers

used

Adjustment of the calibration solutions and the

inner buffer of the electrode.

The instruments can be used for pH or ORP measurements.