Manual OPTISENS AAC 050

OPTISENS AAC 050

26

7.

Output

Main menu

.

.

Basic settings

Basic settings

.

.

Analog output

Analog output

Select output

Selection

.

0 – 20 mA

- between 0/4 – 20 mA

.

Begin 0/4 mA

.

0.00 mg/l

- of the measured value

End 20 mA

Corresponding to 20 mA

4.00 mg/l

Analog output

- between readout of the

Measuring value

Measurement or the

Control values

Serial interface

RS485 (option)

Bus address

Bus address

Nr. 00

Set a number
between 0 and 31
to address the
device via data
bus.

7.1

Current output

You can read out the measured values and the temperature as 0/4…20 mA signals via the current outputs.

With the setting 4…20 mA the resolution is lower, but defective cable connections are immediately evident.
With the parameters Begin and End you define which part of the measuring range you want to read out.

Alternatively you can use the current output as steady-state controller output. In that case assign the current
output to controller S1 or S2, respectively.

7.2

Serial interface RS485 (option)

The devices are available with serial interface RS485 by means of which they can be integrated in a data
bus system. Via the interface, all settings, measured and control values as well as any error messages can
be read out digitally.

Devices with RS485 are automatically delivered with the leaflet „Information on the RS485“ which contains
instructions on the communication and a complete list of the functions available via interface.