4.1

Configuration of the converter

Main menu

.

.

Basic settings

Basic settings

.

.

Configuration

Configuration

Measurement

Temp. Sensor

The instruments AC050 POT provide two measuring inputs.
Input 1 can work with the following sensors:

Parameter

Measuring range

Free Chlorine

Cl2

0.00…4.00 mg/l

Free Chlorine

Cl2

0.00…20.00 mg/l

Chlorine dioxide

ClO2

0.00…4.00 mg/l

Ozone

O3

0.00…4.00 mg/l

Total Chlorine

TCl

0.00…4.00 mg/l

Oxygen

O2 HAZ

0.00…20.00 mg/l

Oxygen

O2 HA

0.000…9.999 mg/l

Hydrogen peroxide

H2O2

0.0…100.0 mg/l

Input 2 works with the temperature sensors Pt100 or NTC.

4.2

Calibration

Main menu

Calibration

DPD Calibration

DPD Calibration

0.48 mg/l

Slope

Enter the concentration determined by
photometric measurement as calibration
value.

025 mV

Calibration of the measurement of Cl2, TCl2, ClO2, O3 and H2O2

1) Switch off the controller. Take a sample of the test water flowing out of the armature and determine the
concentration by photometric DPD measurement or a similar reference method.
2) Enter the determined concentration and calibrate by pressing keys

and : Start with key

and then -

while still applying pressure on this key - additionally press key .
3) Check the displayed slope before switching on the controller.