KROHNE OPTISENS AAC 050 EN User Manual
Page 15
Manual OPTISENS AAC 050
15
OPTISENS AAC 050
4.1
Configuration of the converter
Main menu
.
.
Basic settings
Basic settings
.
.
Configuration
Configuration
Measurement
Temp. Sensor
The instruments AC050 POT provide two measuring inputs.
Input 1 can work with the following sensors:
Parameter
Measuring range
Free Chlorine
Cl2
0.00…4.00 mg/l
Free Chlorine
Cl2
0.00…20.00 mg/l
Chlorine dioxide
ClO2
0.00…4.00 mg/l
Ozone
O3
0.00…4.00 mg/l
Total Chlorine
TCl
0.00…4.00 mg/l
Oxygen
O2 HAZ
0.00…20.00 mg/l
Oxygen
O2 HA
0.000…9.999 mg/l
Hydrogen peroxide
H2O2
0.0…100.0 mg/l
Input 2 works with the temperature sensors Pt100 or NTC.
4.2
Calibration
Main menu
Calibration
DPD Calibration
DPD Calibration
0.48 mg/l
Slope
Enter the concentration determined by
photometric measurement as calibration
value.
025 mV
Calibration of the measurement of Cl2, TCl2, ClO2, O3 and H2O2
1) Switch off the controller. Take a sample of the test water flowing out of the armature and determine the
concentration by photometric DPD measurement or a similar reference method.
2) Enter the determined concentration and calibrate by pressing keys
and : Start with key
and then -
while still applying pressure on this key - additionally press key .
3) Check the displayed slope before switching on the controller.