For an ON/OFF controller you have to set the following parameters:

1) Set points S1 and S2
Set point S1 refers to relay 1, set point S2 refers to relay 2.

2) P range and integral action time for S1 and S2
For an ON/OFF controller set P range = 0 and integral time = 0.

3) Acting direction for S1 and S2
Select “raise“ if the dosage raises the measured value.
Select “lower“ if the dosage lowers the measured value.

4) Optionally a hysteresis
The hysteresis prevents fast switching in the vicinity of the set point. If hysteresis is activated (by setting a
value > 0) the relay switches only when the set point is exceeded by half the hysteresis.

5.2

P / PI controller as impulse-frequency controller

Main menu

.

.

Set points and
controller version

Set points

Set point S1

You can choose different

.

1.50 mg/l

Controller versions

.

P range S1

For S1 and S2

.

0.20 mg/l

Integral time S1

000 sec.

Set point S2

0.50 mg/l

Controller settings

P range S2

Pulse-Freq. S1

0.20 mg/l

0*100/h

Integral time S2

Pulse-Freq. S2

000 sec.

0*100/h

Direction S1

Basic settings

Raise

Basic settings

.

Direction S2

Contr. settings

lower

For an impulse-frequency controller you have to set the following parameters:

1) Set points S1 and S2
S1 refers to relay 1, S2 refers to relay 2.

2) P range and integral action time for S1 and S2
Adjust a P range > 0. For a P controller set integral time = 0, for a PI controller set an integral time > 0.

3) The acting direction for S1 and S2
Select „raise“ if the dosage raises the measured value.
Select „lower“ if the dosage lowers the measured value.

4) Pulse-frequencies for S1 and S2
Enter the maximum pulse-frequency that corresponds to 100% dosage.