beautypg.com

KROHNE OPTISENS AAC 050 EN User Manual

Page 10

background image

Manual OPTISENS AAC 050

OPTISENS AAC 050

10

2.1

Connection diagram panel-mounting converter



Connection

Terminals

Notes
1 = screen
2 = measuring electrode

1- 4

3 = reference electrode

Metal electrodes &
membrane sensors
without electronics

4 = counter electrode


With 2-electrode sensors bridge 3 + 4

1 + 2 = measurement
3 = - 6V

Membrane Sensor with
electronics

1 - 4

4 = + 6V

Pt100/NTC

5 + 6

Analog output

9 + 10

9 = +, 10 = - , max. burden 500 Ohm

Digital input

11 + 12

11 = +, 12 = - , external controller stop or lack-of-water sensor or
flow-meter
Flow measurement: 11 = white, 12 = brown

Relay 1

18 + 19

Relay 2

20 + 21

Relay 3

22 + 23

Alarm Relay

Power supply

24+25+26

check information given on nameplate

RS485 (option)

Sub-D
Sub-D 4/7

3 = +, 8 = -
bridged = terminating impedance activated