KROHNE PROFIFLUX IFS 5000 EN User Manual
Page 4
4
1 Important information for installation: PLEASE NOTE
•
Location and position as required,
but electrode axis X – • – • – • – X
must be approximately horizontal in a
horizontal pipe run.
Y terminal box or converter housing
•
Measuring tube must be completetly filled at all times.
•
Direction of flow is arbitrary. Arrow on flowmeter can normally be ignored.
For exceptions, refer to Section ”Factory settings” in the installation and operating
instructions for the signal converter.
•
Stud bolts and nuts: to fit, make sure there is sufficient room next to the pipe flanges.
•
Vibration: support the pipeline on both sides of the compact flowmeter.
Level of vibration in conformity with IEC 068-2-34:
20-500 Hz, ran dom / 2 g rms / 30 minutes / x, y, z directions.
•
Do not expose to direct sunlight,
fit a sunshade if necessary, not included with flowmeter, to be provided by customer.
X
X
•
Handling
Do not lift flowmeter by the
signal converter housing or
the terminal box.
Do not set flowmeter down on signal
converter housing or terminal box.
•
Use only solventless detergents to clean the signal converter housing (polycarbonate).
•
Temperatures
For operating pressure and vacuum load based on flange standards, see Section 10
“Technical data“.
Y
Ambient temperature
Process temperature
Compact systems
Standard
-25 to +60 °C (-13 to +140 °F)
-60 to +
0
60 °C (-76 to +140 °F)
-25 to +40 °C (-13 to +104 °F)
-60 to +140 °C (-76 to +284 °F)
EEx-version
-25 to +60 °C (-13 to +140 °F)
-20 to +
1
60 °C (- 4 to +140 °F)
-25 to +40 °C (-13 to +104 °F)
-20 to +140 °C (- 4 to +284 °F)
IFS 5000 F (separate): Standard
-25 to +60 °C (-13 to +140 °F)
-60 to +180 °C (-76 to +356 °F)
EEx version
-25 to +60 °C (-13 to +140 °F)
-20 to +
1
60 °C (- 4 to +140 °F)
-25 to +40 °C (-13 to +104 °F)
-20 to +150 °C (- 4 to +302 °F)