1 Important information for installation: PLEASE NOTE

Location and position as required,
but electrode axis X – • – • – • – X
must be approximately horizontal in a
horizontal pipe run.

Y terminal box or converter housing

Measuring tube must be completetly filled at all times.

Direction of flow is arbitrary. Arrow on flowmeter can normally be ignored.
For exceptions, refer to Section ”Factory settings” in the installation and operating
instructions for the signal converter.

Stud bolts and nuts: to fit, make sure there is sufficient room next to the pipe flanges.

Vibration: support the pipeline on both sides of the compact flowmeter.
Level of vibration in conformity with IEC 068-2-34:
20-500 Hz, ran dom / 2 g rms / 30 minutes / x, y, z directions.

Do not expose to direct sunlight,
fit a sunshade if necessary, not included with flowmeter, to be provided by customer.

X

X

Handling
Do not lift flowmeter by the
signal converter housing or
the terminal box.

Do not set flowmeter down on signal
converter housing or terminal box.

Use only solventless detergents to clean the signal converter housing (polycarbonate).

Temperatures
For operating pressure and vacuum load based on flange standards, see Section 10
“Technical data“.

Y

Ambient temperature

Process temperature

Compact systems

Standard

-25 to +60 °C (-13 to +140 °F)

-60 to +

0

60 °C (-76 to +140 °F)

-25 to +40 °C (-13 to +104 °F)

-60 to +140 °C (-76 to +284 °F)

EEx-version

-25 to +60 °C (-13 to +140 °F)

-20 to +

1

60 °C (- 4 to +140 °F)

-25 to +40 °C (-13 to +104 °F)

-20 to +140 °C (- 4 to +284 °F)

IFS 5000 F (separate): Standard

-25 to +60 °C (-13 to +140 °F)

-60 to +180 °C (-76 to +356 °F)

EEx version

-25 to +60 °C (-13 to +140 °F)

-20 to +

1

60 °C (- 4 to +140 °F)

-25 to +40 °C (-13 to +104 °F)

-20 to +150 °C (- 4 to +302 °F)