Operation, 3 status messages and diagnostic information – KROHNE SMARTSENS PH 8320 EN User Manual
Page 23
OPERATION
4
23
SMARTSENS PH 8320
www.krohne.com
11/2013 - 4002543902 - MA SMARTSENS PH 8320 R02 en
4.3 Status messages and diagnostic information
Measurements out of specification
Failure
Maintenance
Message
Message
Message
Message
Description
Description
Description
Description
Action
Action
Action
Action
S01
pH value > pH max
pH value out of range
Consider the measuring range limits, otherwise
select a suitable sensor for the process conditions of
the application.
S02
pH value < pH min
pH value out of range
S03
Temp. value > Temp max
Temperature value out of range
S04
Temp. value < Temp min
Temperature value out of range
Message
Message
Message
Message
Description
Description
Description
Description
Action
Action
Action
Action
F01
Glass resistance > max. MOhm
Replace the sensor
F02
Glass resistance < min. MOhm
Message
Message
Message
Message
Description
Description
Description
Description
Action
Action
Action
Action
M01
Slope < -65 mV/pH
Recalibrate the sensor. For more information refer
to
Calibration with PACTwareTM on page 17
M02
Slope > -50 mV/pH
M03
Zero point < -99 mV
M04
Zero point > +99 mV
M05
Maintenance interval expired
.book Page 23 Monday, November 18, 2013 10:42 AM