beautypg.com

Operation – KROHNE SMARTSENS PH 8320 EN User Manual

Page 19

background image

OPERATION

4

19

SMARTSENS PH 8320

www.krohne.com

11/2013 - 4002543902 - MA SMARTSENS PH 8320 R02 en

• Press Next

Next

Next

Next to proceed.

• Select no

no

no

no to deactivate the function manual hold

manual hold

manual hold

manual hold.

• Press Next

Next

Next

Next to leave the calibration menu.

In a fully functional sensor, the optimal slope is -59 mV for each pH unit and the optimal zero
point is 0 mV at pH 7. The slope should at least have a value between -50...-65 mV per pH unit.
Re-calibrate the sensor if the slope does not approximate those limits. The pH sensor ages, the
slope gets flatter and the zero error increases.

4.1.2 Calibration with HART

®

Handheld 475 FIELD COMMUNICATOR

Manual hold

• Start the function calibration

calibration

calibration

calibration in menu mode Quick Setup

Quick Setup

Quick Setup

Quick Setup or Setup

Setup

Setup

Setup.

• Activate the function manual hold

manual hold

manual hold

manual hold to avoid an alarm.

• Select Yes

Yes

Yes

Yes.

• Press Enter

Enter

Enter

Enter to proceed.

Temperature compensation

• Set the temperature compensation

temperature compensation

temperature compensation

temperature compensation for calibration.

• Select between automatic

automatic

automatic

automatic or manual

manual

manual

manual.

i

If you have selected manual temperature compensation

manual temperature compensation

manual temperature compensation

manual temperature compensation, please enter the temperature of

the pH buffer solution. Default setting: 25°C / 77°F.

• Press Enter

Enter

Enter

Enter to proceed.

Start calibration procedure

• Dip the sensor into the first pH buffer solution 1 and wait till the value is stable.
• Press Ok

Ok

Ok

Ok to proceed.

i

The measurement of pH buffer solution 1

pH buffer solution 1

pH buffer solution 1

pH buffer solution 1 starts and the values pH and temperature appear.

After approx. 10 seconds the following values appear:
pH value
temperature value
Status buffer detection → Finished
Value indentified buffer

• Press Ok

Ok

Ok

Ok to confirm the identified pH buffer solution. If neccessary change the value of the pH

buffer solution manually.

• Press Enter

Enter

Enter

Enter to proceed.

i

If the sensor does not recognise the pH buffer solution the messages
Identified buffer

Identified buffer

Identified buffer

Identified buffer → 0 pH

0 pH

0 pH

0 pH and Buffer not recognised

Buffer not recognised

Buffer not recognised

Buffer not recognised appear.

• Press Ok

Ok

Ok

Ok to proceed.

i

Enter the value of the pH buffer solution manually.

CAUTION!

Never touch or scratch the pH sensitive glass tip of the sensor.

Make sure that the sensor tip is clean and dust-free. If necessary, clean the tip as described
on page 24.

CAUTION!
Moisture inside the sensor connector must be avoided! Moisture may cause a shortcut and
deliver erratic readings!
If moisture has entered the connector dry it with air (e.g. hair blower).

.book Page 19 Monday, November 18, 2013 10:42 AM