Operation – KROHNE SMARTSENS PH 8320 EN User Manual

Page 18

background image

4

OPERATION

18

SMARTSENS PH 8320

www.krohne.com

11/2013 - 4002543902 - MA SMARTSENS PH 8320 R02 en

• Dip the sensor into the second pH buffer solution 3 and wait till the value ist stable.
• Press Next

Next

Next

Next to proceed.

i

The measurement of pH buffer solution 2

pH buffer solution 2

pH buffer solution 2

pH buffer solution 2 starts and the values pH and temperature appear.

After approx. 10 seconds the following values appear:
pH value
temperature value
Indentified buffer value

• Press Next

Next

Next

Next to confirm the identified pH buffer solution. If neccessary change the value of the

pH buffer solution manually.
If the sensor does not recognise the pH buffer solution the message buffer not recognised

buffer not recognised

buffer not recognised

buffer not recognised

appears.

• Press Next

Next

Next

Next to proceed.

i

The value 0 for the pH buffer solution appears. Enter the value of the pH buffer solution
manually.

• Press Next

Next

Next

Next to proceed with the calibration procedure.

• Clean the sensor with demineralised water 2.

i

After a successful calibration the following values appear:
Buffer solution 1
Buffer solution 2
Offset in mV
Slope in mV/pH

• Press Next

Next

Next

Next to proceed.

• Set the calibration date with DD-MM-YYYY and calibration time with HH-MM-SS
• Press Next

Next

Next

Next to proceed.

i

The message Save values?

Save values?

Save values?

Save values? appears.

• Select yes

yes

yes

yes to save the values.

• Press Next

Next

Next

Next to proceed.

• Select no

no

no

no to deactivate the function manual hold

manual hold

manual hold

manual hold.

• Press Next

Next

Next

Next to leave the calibration menu.

A pH calibration is necessary in regular intervals or when installing a new pH sensor.

If the calibration was not successful, the message Warning! Offset out of range!

Warning! Offset out of range!

Warning! Offset out of range!

Warning! Offset out of range! appears behind

the value Offset

Offset

Offset

Offset and/or the message Warning! Slope out of range!

Warning! Slope out of range!

Warning! Slope out of range!

Warning! Slope out of range! behind the value Slope

Slope

Slope

Slope.

In case the value of the slope is not plausible, addionally the message No saving possible

No saving possible

No saving possible

No saving possible

appears.

• Press Next

Next

Next

Next to proceed.

i

The messages Warning! Offset out of range!

Warning! Offset out of range!

Warning! Offset out of range!

Warning! Offset out of range! or Warning! Slope out of range!

Warning! Slope out of range!

Warning! Slope out of range!

Warning! Slope out of range! appear.

• Press Next

Next

Next

Next to proceed.

• Set the calibration date with DD-MM-YYYY and calibration time with HH-MM-SS
• Press Next

Next

Next

Next to proceed.

i

The message Save values?

Save values?

Save values?

Save values? appears.

• Select between yes

yes

yes

yes or no

no

no

no.

• Press Next

Next

Next

Next to proceed.

• Select no

no

no

no to deactivate the function manual hold

manual hold

manual hold

manual hold.

• Press Next

Next

Next

Next to leave the calibration menu.

If the message Warning! Slope out of range!

Warning! Slope out of range!

Warning! Slope out of range!

Warning! Slope out of range! appears together with the message No saving

No saving

No saving

No saving

possible

possible

possible

possible, no data can be saved.