Operation – KROHNE OPTISENS PH 8500 EN User Manual
Page 35
OPERATION
4
35
OPTISENS PH 8500
www.krohne.com
04/2012 - 4001927601 - MA OPTISENS PH 8500 R01 en
Step 3a: accessing the calibration menu via the main menu setup
setup
setup
setup
• You can start the calibration procedure now as described in "Step 4".
Step 3b: accessing the calibration menu via the main menu quick setup
quick setup
quick setup
quick setup
• You can start the calibration procedure now as described in "Step 4".
•
Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu quick setup
quick setup
quick setup
quick setup is highlighted.
•
Press
or
until the main menu setup
setup
setup
setup is highlighted.
MAIN MENU
A quick setup
B test
> C setup
> C setup
> C setup
> C setup
D service
•
Press >
>>> to enter the chosen menu.
You are on the first submenu level. In the upper line of the display "setup" and "c1"
appears, beneath the submenu process input A
process input A
process input A
process input A is highlighted.
•
Press
or
to select process input A
process input A
process input A
process input A or process input B
process input B
process input B
process input B is highlighted.
Choose process input A or B where ever pH is configured.
•
Press >>>> to enter the chosen menu.
You are on the second submenu level. In the upper line of the
display "process input A" and "C1.1" appears, beneath the
submenu parameter
parameter
parameter
parameter is highlighted.
•
Press
or
until the submenu pH cal.
pH cal.
pH cal.
pH cal. is highlighted.
•
Press >>>> to enter the chosen menu.
•
Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu quick setup
quick setup
quick setup
quick setup is highlighted.
MAIN MENU
> A quick setup
> A quick setup
> A quick setup
> A quick setup
B test
C setup
D service
•
Press >>>> to enter the chosen menu.
You are on the first submenu level. In the upper line of the display "quick
setup" and "a" appears, beneath the submenu language
language
language
language is highlighted.
•
Press
or
until the submenu pH cal. A
pH cal. A
pH cal. A
pH cal. A or B
B
B
B is highlighted, wherever pH
is configurated
•
Press >>>> to enter the chosen menu.
.book Page 35 Wednesday, July 4, 2012 9:52 AM