Operation, 3 calibration – KROHNE OPTISENS PH 8500 EN User Manual
Page 29
OPERATION
4
29
OPTISENS PH 8500
www.krohne.com
04/2012 - 4001927601 - MA OPTISENS PH 8500 R01 en
4.3 Calibration
4.3.1 Temperature compensation
There are three basic options for temperature compensation:
• automatic:
automatic:
automatic:
automatic: the signal converter will automatically compensate temperature influences using
the information of a Pt100 or Pt1000 temperature sensor.
• manual:
manual:
manual:
manual: the signal converter will compensate temperature influences using a manually
entered value; this option only makes sense if the temperature of the measured medium is
quite constant.
• off:
off:
off:
off: temperature compensation is disabled.
The menu for the type of temperature compensation offers the following options:
• Pt1000
Pt1000
Pt1000
Pt1000: choose this option if there is an external Pt1000 temperature sensor connected to the
signal converter.
• Pt100
Pt100
Pt100
Pt100: choose this option if the electrode has an integrated Pt100 temperature measurement
or if there is an external Pt100 temperature sensor connected to the signal converter.
• manual
manual
manual
manual: choose this option if there is no internal or external temperature sensor connected
to the signal converter.
• process input A
process input A
process input A
process input A: choose this option if there is a temperature sensor connected to input A of
the signal converter.
INFORMATION!
If you choose no compensation, the measured pH value will most likely deviate from the actual
pH value. The reason is that the pH value of a specific medium varies depending on the
temperature of the medium.
INFORMATION!
In the basic configuration the setting of the temperature compensation is Pt100. If you use a pH
sensor without Pt100/1000 than select the temperature compensation "manual" and enter the
value. Otherwise an error will be displayed.
.book Page 29 Wednesday, July 4, 2012 9:52 AM