beautypg.com

Operation, 3 calibration – KROHNE OPTISENS PH 8500 EN User Manual

Page 29

background image

OPERATION

4

29

OPTISENS PH 8500

www.krohne.com

04/2012 - 4001927601 - MA OPTISENS PH 8500 R01 en

4.3 Calibration

4.3.1 Temperature compensation

There are three basic options for temperature compensation:

• automatic:

automatic:

automatic:

automatic: the signal converter will automatically compensate temperature influences using
the information of a Pt100 or Pt1000 temperature sensor.

• manual:

manual:

manual:

manual: the signal converter will compensate temperature influences using a manually
entered value; this option only makes sense if the temperature of the measured medium is
quite constant.

• off:

off:

off:

off: temperature compensation is disabled.

The menu for the type of temperature compensation offers the following options:

• Pt1000

Pt1000

Pt1000

Pt1000: choose this option if there is an external Pt1000 temperature sensor connected to the
signal converter.

• Pt100

Pt100

Pt100

Pt100: choose this option if the electrode has an integrated Pt100 temperature measurement
or if there is an external Pt100 temperature sensor connected to the signal converter.

• manual

manual

manual

manual: choose this option if there is no internal or external temperature sensor connected
to the signal converter.

• process input A

process input A

process input A

process input A: choose this option if there is a temperature sensor connected to input A of
the signal converter.

INFORMATION!
If you choose no compensation, the measured pH value will most likely deviate from the actual
pH value. The reason is that the pH value of a specific medium varies depending on the
temperature of the medium.

INFORMATION!
In the basic configuration the setting of the temperature compensation is Pt100. If you use a pH
sensor without Pt100/1000 than select the temperature compensation "manual" and enter the
value. Otherwise an error will be displayed.

.book Page 29 Wednesday, July 4, 2012 9:52 AM