4.2.2 Menu B, test
The procedure to start the simulation process is the same for all functions:
• Choose the function with the help of ↓ or ↑ and press ^.
i
You see the two options "set value" (opens the editor to enter the simulation value) and
"break" (exits the menu without simulation).
• Choose the desired option with the help of ↑ or ↓ and press ^.
i
If you chose "set value", the device asks "start simulation" and offers the options "no" (exits
the menu without simulation) or "yes" (starts the simulation finally).
• Choose the desired option with the help of ↑ or ↓ and press ^.
i
If you chose "yes", the simulation starts.
B1,sim.process inp.A
B2,sim.process inp.B
B4, actual values
INFORMATION!
Note that the appearance of some sub-menus depends on the hardware setting and the used
sensor(s). Also only the sensor relevant menus and sub-menus are shown here in detail. For all
other menu functions refer to the MAC 100 signal converter manual.
Level
Designation / function
Settings / descriptions
B1.1
B1.1
B1.1
B1.1
B2.1
B2.1
B2.1
B2.1
temperature
In this menu the temperature can be simulated.
B1.5
B1.5
B1.5
B1.5
B2.5
B2.5
B2.5
B2.5
ORP
ONLY if switched to ORP!
ONLY if switched to ORP!
ONLY if switched to ORP!
ONLY if switched to ORP!
(for dual channel version
only; refer to pH/ORP
sensor manual for further
information)
In this menu the relative concentration of ORP can be simulated. For dual
channel version only; refer to pH/ORP sensor manual for further
information. ONLY if switched to ORP!
ONLY if switched to ORP!
ONLY if switched to ORP!
ONLY if switched to ORP!
B1.7
B1.7
B1.7
B1.7
B2.7
B2.7
B2.7
B2.7
pH
(for dual channel version
only; refer to pH/ORP
sensor manual for further
information)
In this menu the relative concentration of pH can be simulated. For dual
channel version only; refer to pH/ORP sensor manual for further
information.
Level
Designation / function
Settings / descriptions
This menu groups several functions which allow to display the corresponding actual reading. The shown measurements
are depending on the device configuration.
B4.1
B4.1
B4.1
B4.1
operating hours
This menu shows the operating time of the devices in hours.
B4.2
B4.2
B4.2
B4.2
process input A
In this menu the measurements from process input A can be read.
B4.3
B4.3
B4.3
B4.3
process input B
In this menu the measurements from process input B can be read.For dual
channel version only; refer to pH/ORP sensor manual for further
information.