Operation – KROHNE OPTISENS PH 8500 EN User Manual
Page 28
4
OPERATION
28
OPTISENS PH 8500
www.krohne.com
04/2012 - 4001927601 - MA OPTISENS PH 8500 R01 en
C1.15.5
C2.15.5
temp. comp.
Menu for activating the temperature compensation parameters for the
measurement.
Options:
•
Linear: linear temperature compensation.
•
Off: temperature compensation is disabled.
C1.16
C1.16
C1.16
C1.16
C2.16
C2.16
C2.16
C2.16
ORP cal.
menu item for calibrating the ORP sensor
C1.16.1
C2.16.1
prepare calibration
view actual ORP value
C1.16.2
C2.16.2
ref. value ORP
Enter the reference value of the probe in mV
C1.16.3
C2.16.3
start calibration?
Start calibration procedure.
C1.16.4
C2.16.4
stored value
View stored value of calibration.
C1.17
C1.17
C1.17
C1.17
C2.17
C2.17
C2.17
C2.17
pH cal.
menu item for calibration the pH sensor
C1.17.1
C2.17.1
temp.comp
Menu for activating the temperature compensation parameters for the
calibration
Options:
•
off: temperature measurement is disabled
•
manual: temperature value has to be entered manually
•
automatic: temperature measurement is performed as configured
C1.17.3
C2.17.3
calib. buffer 1
Start calibration procedure in buffer 1
C1.17.4
C2.17.4
pH 1 set value
Enter the reference value of the probe in pH
C1.17.6
C2.17.6
calib. buffer 2
Start calibration procedure in buffer 2
C1.17.7
C2.17.7
pH 2 set value
Enter the reference value of the probe in pH
CAUTION!
If you choose for measurement the temperature compensation "linear" than choose between
"automatic" or "manual" for the temperature compensation during calibration. If you choose for
measurement the temperature compensation "off" than choose also "off" for the temperature
compensation during calibration.
Level
Designation / function
Settings / descriptions
.book Page 28 Wednesday, July 4, 2012 9:52 AM