beautypg.com

Operation, 4 troubleshooting – KROHNE OPTISENS PH 8300 EN User Manual

Page 38

background image

4

OPERATION

38

OPTISENS PH 8300

www.krohne.com

04/2012 - 4001926701 - MA OPTISENS PH 8300 R01 en

4.3.3 Calibration log

Accessing the calibtration log

• You are on the data level and you see the calibration history. With the help of

or

you can

scroll through the different entries.

• If you want to return to the measuring mode press ^

^

^

^ several times until you reach this mode.

4.4 Troubleshooting

INFORMATION!
In order to show the history of the calibrations, the converter has a calibration logbook function.
Up to 64 entries of the calibration history are stored including date and time.

Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu quick setup

quick setup

quick setup

quick setup is highlighted.

Press

or

until the main menu test

test

test

test is highlighted.

MAIN MENU

A quick setup

> B test

> B test

> B test

> B test

C setup

D service

Press >>>> to enter the chosen menu.

You are on the first submenu level. In the upper line of the display "test" and "B1" appears,

beneath the submenu sim.process input A

sim.process input A

sim.process input A

sim.process input A is highlighted.

Press or until the submenu logbooks

logbooks

logbooks

logbooks is highlighted.

Press >>>> to enter the chosen menu.

You are on the second submenu level. In the upper line of the display "logbooks" and

"B1" appears, beneath the submenu status log

status log

status log

status log is highlighted.

Press

or

until the submenu calibration log

calibration log

calibration log

calibration log is highlighted.

Press >

>>> to enter the chosen menu.

Problem

Possible cause

Remedy

The pH electrode

does not deliver a

signal.

Mechanical damage of the

glass bulb, e.g. small

cracks. This will shortcut

the ion exchange and

deliver a pH7 reading.

Exchange electrode.

Moisture inside the

electrode connector. This

will shortcut the signal

between pH glass half cell

and reference half cell

and deliver a reading of

pH7.

Clean the connector (electrode/cable) with pure water

and dry with air (e.g. hair blower).

The pH electrode

delivers an

unstable signal.

The diaphragm in the

reference half cell does

not provide good contact

to the process medium

due to drying up or

coatings.

Clean the diaphragm with hot soap or acid using a
soft tissue (details on page 39).

Submerge electrode in water and increase the
temperature to 50...60°C / 122...140°F.

Submerge electrode in 3 molar KCl solution at
ambient temperature. The decrease in temperature
will cause the reference half cell to suck in KCl
solution through the diaphragm and regenerate the
diaphragms functionality.

.book Page 38 Tuesday, July 3, 2012 2:56 PM