INSTALLATION

3

13

OPTISENS PH 8300

www.krohne.com

04/2012 - 4001926701 - MA OPTISENS PH 8300 R01 en

3.4 Pre-installation requirements

Unpacking the electrode

• Loosen the storage cap which is screwed or/and pushed on to the plastic tube 1.
• Gently pull the electrode out of the plastic tube 2.
• Lay the electrode on a soft mat/tissue 3.
• Screw or push the provided sealing cap on to the plastic tube, using O-ring as shown in the

drawing 4. Keep the storage cap (the one with the hole in it) in the original packaging.

3.5 Electrical connection

CAUTION!

Never touch or scratch the pH sensitive glass tip of the electrode.

Make sure that the glass tip is clean and dust-free. If necessary, clean the tip as described on
page 39

Figure 3-2: Handling the electrode

DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!

DANGER!
Observe the national regulations for electrical installations!

WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the
electrical components of the measuring device may only be carried out by properly trained
specialists.

INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.