Operation – KROHNE OPTISENS PH 8300 EN User Manual
Page 27
OPERATION
4
27
OPTISENS PH 8300
www.krohne.com
04/2012 - 4001926701 - MA OPTISENS PH 8300 R01 en
4.2.3 Menu C, setup
C1, process input A
C1, process input A
C1, process input A
C1, process input A
C2, process input B
C2, process input B
C2, process input B
C2, process input B
INFORMATION!
The signal converter has a dual process input, A and B. Each process input has an own submenu
in this main menu. Process input A is always present, i.e. there is always a board in the interface
"Pos.A" in the connection area. The interface of process input B only has a board with the dual
channel signal converter. Be aware that the definition which kind of measurement a process
input can do is defined when ordering the device. The configuration cannot be changed later.
INFORMATION!
Note that the appearance of some submenus depends on the hardware setting and the used
sensor(s).
Level
Designation / function
Settings / descriptions
Process input A and B can be either a sensor 1 or a sensor 2. Further information about the type of sensor 1 or 2 please
refer to MAC 100 manual "Sensor input combinations". Process input A is always present, process Input B can be
present.
Note: The exchange of a sensors 1 with a sensor 2, or vice versa, can only be done by the manufacturer!
Depending on the sensor which is connected to a slot A or B the menu changes.
C1.1
C1.1
C1.1
C1.1
C2.1
C2.1
C2.1
C2.1
parameter pH
parameter ORP
This menu item is for selecting the probe which is connected to process
input A/B. The entries of this selection depends on the chosen device
configuration. The device configuration is customer specific and set during
production.
C1.7
C1.7
C1.7
C1.7
C2.7
C2.7
C2.7
C2.7
inner buffer
setting for inner buffer of the pH sensor
C1.8
C1.8
C1.8
C1.8
C2.8
C2.8
C2.8
C2.8
zero point
this menu item shows the actual zero point to the calibrated pH/ORP sensor
C1.9
C1.9
C1.9
C1.9
C2.9
C2.9
C2.9
C2.9
slope
this menu item shows the actual slope point to the calibrated pH/ORP
sensor
C1.10
C1.10
C1.10
C1.10
C2.10
C2.10
C2.10
C2.10
calibration buffer
this menu item shows the used buffer concentration for the last calibration
C1.15
C1.15
C1.15
C1.15
C2.15
C2.15
C2.15
C2.15
temperature
Menu for temperature measurement. Available for sensor 1 and sensor 2.
C1.15.1
C2.15.1
probe
Options:
•
manual: used if no internal or external temperature sensor is connected to
the signal converter
•
Pt100: used if the pH electrode has an integrated Pt temperature
measurement or if an external Pt100 temperature sensor is connected to
the signal converter
•
Pt1000: used if an external Pt1000 temperature sensor is connected to the
signal converter
•
process input A: choose this option if there is a temperature sensor
connected to input A of the signal converter
C1.15.2
C2.15.2
manual
Only available if C1.15.1 or C2.15.1 is set to "manual".
C1.15.3
C2.15.3
correction
Offset correction for temperature measurement. Not available if C1.15.1 or
C2.15.1 is set to "manual".
C1.15.4
C2.15.4
limitation
Measuring ranges for temperature measurement.
.book Page 27 Tuesday, July 3, 2012 2:56 PM