Operation – KROHNE OPTISENS PH 8300 EN User Manual

Page 27

background image

OPERATION

4

27

OPTISENS PH 8300

www.krohne.com

04/2012 - 4001926701 - MA OPTISENS PH 8300 R01 en

4.2.3 Menu C, setup

C1, process input A

C1, process input A

C1, process input A

C1, process input A
C2, process input B

C2, process input B

C2, process input B

C2, process input B

INFORMATION!
The signal converter has a dual process input, A and B. Each process input has an own submenu
in this main menu. Process input A is always present, i.e. there is always a board in the interface
"Pos.A" in the connection area. The interface of process input B only has a board with the dual
channel signal converter. Be aware that the definition which kind of measurement a process
input can do is defined when ordering the device. The configuration cannot be changed later.

INFORMATION!
Note that the appearance of some submenus depends on the hardware setting and the used
sensor(s).

Level

Designation / function

Settings / descriptions

Process input A and B can be either a sensor 1 or a sensor 2. Further information about the type of sensor 1 or 2 please

refer to MAC 100 manual "Sensor input combinations". Process input A is always present, process Input B can be

present.

Note: The exchange of a sensors 1 with a sensor 2, or vice versa, can only be done by the manufacturer!

Depending on the sensor which is connected to a slot A or B the menu changes.

C1.1

C1.1

C1.1

C1.1

C2.1

C2.1

C2.1

C2.1

parameter pH

parameter ORP

This menu item is for selecting the probe which is connected to process

input A/B. The entries of this selection depends on the chosen device

configuration. The device configuration is customer specific and set during

production.

C1.7

C1.7

C1.7

C1.7

C2.7

C2.7

C2.7

C2.7

inner buffer

setting for inner buffer of the pH sensor

C1.8

C1.8

C1.8

C1.8

C2.8

C2.8

C2.8

C2.8

zero point

this menu item shows the actual zero point to the calibrated pH/ORP sensor

C1.9

C1.9

C1.9

C1.9

C2.9

C2.9

C2.9

C2.9

slope

this menu item shows the actual slope point to the calibrated pH/ORP

sensor

C1.10

C1.10

C1.10

C1.10

C2.10

C2.10

C2.10

C2.10

calibration buffer

this menu item shows the used buffer concentration for the last calibration

C1.15

C1.15

C1.15

C1.15

C2.15

C2.15

C2.15

C2.15

temperature

Menu for temperature measurement. Available for sensor 1 and sensor 2.

C1.15.1

C2.15.1

probe

Options:

manual: used if no internal or external temperature sensor is connected to
the signal converter

Pt100: used if the pH electrode has an integrated Pt temperature
measurement or if an external Pt100 temperature sensor is connected to
the signal converter

Pt1000: used if an external Pt1000 temperature sensor is connected to the
signal converter

process input A: choose this option if there is a temperature sensor
connected to input A of the signal converter

C1.15.2

C2.15.2

manual

Only available if C1.15.1 or C2.15.1 is set to "manual".

C1.15.3

C2.15.3

correction

Offset correction for temperature measurement. Not available if C1.15.1 or

C2.15.1 is set to "manual".

C1.15.4

C2.15.4

limitation

Measuring ranges for temperature measurement.

.book Page 27 Tuesday, July 3, 2012 2:56 PM