INSTALLATION

3

19

OPTISENS PH 8300

www.krohne.com

04/2012 - 4001926701 - MA OPTISENS PH 8300 R01 en

3.7.3 Mounting to SENSOFIT IMM 1000 immersion holder

INFORMATION!
The immersion assembly is an optional accessory and not part of the standard scope of delivery.

CAUTION!

Never touch or scratch the pH sensitive glass tip of the electrode.

Make sure that the glass tip is clean and dust-free. If necessary, clean the tip as described on
page 39

Figure 3-10: Overview of the immersion holder

1 Immersion holder
2 Protective cage
3 Spigot nut
4 Clamps
5 Electrode
6 Cap with cable gland
7 Sensor cable
8 Sensor connector

CAUTION!
Moisture inside the sensor connector must be avoided! Moisture will shortcut the mV signal
between the electrodes and deliver erratic readings!
If moisture has entered the connector dry it with air (e.g. hair blower).