DEVICE DESCRIPTION

2

9

OPTITEMP TRA/TCA

www.krohne.com

10/2010 - 4001087001 - Ex AD OPTITEMP TRA/TCA R02 en

2.6 Protection types

2.6.1 Protection type with zone separator for gas hazard areas

In protection type "intrinsic safety" the industrial thermometer meets the requirements of
EN 60079-11. The explosion protection is guaranteed because the current and voltage are
limited so that ignitable energy cannot occur.

If zone 0 and zone 1 are separated as per EN 60079-26, this will ensure that a thermowell is
fitted properly. In such cases the thermometer marking is:

II 1/2G Ex ia IIC T6

II 1/2G Ex ia IIC T6

II 1/2G Ex ia IIC T6

II 1/2G Ex ia IIC T6

Components of the Ex marking and what they mean

2.6.2 Type of protection for gas hazard areas

In protection type "intrinsic safety" the industrial thermometer meets the requirements of
EN 60079-11. The explosion protection is guaranteed because the current and voltage are
limited so that ignitable energy cannot occur. In such cases the thermometer marking is:

II 1G Ex ia IIC T6

II 1G Ex ia IIC T6

II 1G Ex ia IIC T6

II 1G Ex ia IIC T6

Components of the Ex marking and what they mean

II

IIII

II

Group II explosion protection

1/2

1/2

1/2

1/2

Equipment category 1/2

G

G

G

G

Gas explosion protection

Ex ia

Ex ia

Ex ia

Ex ia

Intrinsically safe equipment

IIC

IIC

IIC

IIC

Gas groups IIA, IIB, IIC

T6

T6

T6

T6

Temperature classes T6...T1

II

IIII

II

Group II explosion protection

1111

Equipment category 1

G

G

G

G

Gas explosion protection

Ex ia

Ex ia

Ex ia

Ex ia

Intrinsically safe equipment

IIC

IIC

IIC

IIC

Gas groups IIA, IIB, IIC

T6

T6

T6

T6

Temperature classes T6...T1